|
DRAWOVER ANAESTHESIA REVIEW
Dr Scott Simpson, FANZCA, FFPMANZCA, Staff
Anaesthetist, Townsville Hospital, PO Box 670, Queensland, Australia 4810.,
e-mail: scott.simpson@health.qld.gov.au and Dr Iain Wilson FRCA, Royal
Devon and Exeter Healthcare NHS Trust, Barrack Road, Exeter, Devon. UK
This paper discusses the equipment and role of
drawover anaesthesia for the uninitiated ‘plenum anaesthetist’
while also delving into some of the finer aspects for experienced users.
Overview
Drawover anaesthesia is simple in both concept
and delivery, has stood the tests of time, and travels well. The equipment
is generally robust, versatile, easily maintained, and relatively inexpensive.
Why, then, is it not more popular? Possible explanations are included
in table 1.
History
Since the introduction of open drop volatile
anaesthesia with ether, and later chloroform, anaesthetists have sought
to refine vapour delivery in response to a variety of different clinical
goals with different volatile agents. From an historical perspective,
drawover and continuous flow (plenum) anaesthesia have been developed
in parallel, ever since vaporisers first began to replace open drop methods
at the turn of the 20th century.
![[Top]](../graphics/top_bult.gif)
What is drawover anaesthesia?
It is simply the act of drawing a carrier gas
over a volatile liquid for the purpose of adding the vapour from that
liquid to the carrier gas. This carrier gas/vapour mixture is then directed
to the patient by a ‘circuit’. In drawover systems the carrier
gas is drawn through the vaporiser either by the patient’s own
respiratory efforts, or by a self-inflating bag or manual bellows with
a one-way valve placed downstream from the vaporiser. Drawover systems
operate at less than, or at ambient pressure, and flow through
the system is ‘intermittent’, varying with different phases
of inspiration, and ceasing in expiration.
A one-way valve prevents reverse flow in the circuit.
This is different to plenum (Latin derivative,
opposite to vacuum) anaesthesia in which a carrier gas is pushed through
the vaporiser at a constant rate. In plenum circuits the anaesthetic is
then collected in a circuit with a reservoir bag or bellows. Pressure
fluctuations in the circuit caused by patient respiration, whether spontaneous
or applied, do not involve or affect the vaporiser. Plenum systems operate
at higher than ambient pressure.
The basic draw-over system is shown in figure
1.
Practical significance
Draw over systems are simple to assemble and
use, and can operate without fresh gas supplies. They are lightweight
and portable. Plenum systems are more technically complex, and need a
well-regulated, constant, positive pressure gas supply. They require a
more sophisticated anaesthetic ‘machine’ to support them.
The transport of gas cylinders for plenum systems is both expensive
and potentially hazardous. Therefore drawover systems have obvious advantages
in remote locations, in under-resourced countries, and in ‘field/military’
anaesthesia.
| Table
1 Drawover anaesthesia |
|
Advantages
- Simplicity of concept and assembly,
with inherent safety
- No need for pressurised gas supply,
regulators and flow meters
- Minimum FiO2
is ~21%
- Robust, reliable, easily serviced
equipment
- Low cost (purchase and maintenance)
- Portable, suitable for field anaesthesia
|
Disadvantages
- Decreasing familiarity with the technique
and equipment
- Vaporiser limitations
- Filling systems not agent specific
(potential advantage)
- Basic temperature compensation, affecting
performance at extremes
- Less easy to observe spontaneous ventilation
with self inflating bag
- Cumbersome in paediatric use, unless
lightweight tubing is available
|
![[Top]](../graphics/top_bult.gif)
Supplemental oxygen?
In principle the 21% oxygen in air is diluted
by the addition of vapour in the vaporiser, allowing a potentially ‘hypoxic
mixture’ to be delivered to the patient. This is a theoretical
issue, rather than a practical one, as the vapour concentration is small,
and it is unlikely that the FiO 2
falls below 18%, the international
standard for oxygen analyser alarms. It is more
important to consider the respiratory physiological effects of general
anaesthesia which tend to reduce ventilation and increase shunting of
blood within the lung (V/Q mismatch). Therefore hypoxia becomes a clinical
problem when using halothane or isoflurane with spontaneous ventilation
(SV) in air, and supplemental oxygen is necessary 1,2
. The problem is reduced, but abolished,
when applying intermittent positive pressure
ventilation (IPPV). Ether can be delivered in air (without supplemental
oxygen), in IPPV mode, presumably because it causes less intrapulmonary
shunting and tends to stimulate ventilation, rather than depress it. When
used in unsupplemented air with spontaneous respiration, some patients
will desaturate.
In draw over systems supplemental oxygen is administered
via a T-piece connection mounted to the intake port of the vaporiser.
To maximise the inspired oxygen concentration (FiO 2
) a ‘reservoir tube’
is attached to the T-piece, as shown in figure 1. A 1m length of corrugated
tubing with an internal volume of 415ml allows an FiO 2
of at least 30% with an oxygen flow rate
of 1.0 litre/min, and 60% at 4 litres/min, at normal adult minute ventilation
3 .
With higher minute volumes the FiO 2
falls due to increased air dilution;
at lower minute volumes the FiO 2
is higher. Oxygen may be sourced from
cylinders, or an oxygen concentrator. 4,5,6
The Houtonox oxygen flow control device is a
simple, single stage reducing valve (regulator) that is suitably pin-indexed
to fit directly to an oxygen cylinder. Adaptors for bull-nose connections
are available. Flow rates that can be set are fixed at 1 and 4 litres/min,
which are ideal in combination with draw over systems 7
. The device is accurate and sturdy, and
allows maximum benefit from limited oxygen supplies.
![[Top]](../graphics/top_bult.gif)
Equipment used in Drawover Anaesthesia
Vaporisers
The ideal drawover vaporiser needs to have low
internal resistance to gas flow to allow easy spontaneous ventilation,
while vapour output should be constant for a given dial setting over a
wide range of minute volumes and ambient temperatures. Other desirable
qualities are that the circuit connectors comply with international standards
and that chamber filling is visible. These requirements determine careful
vapour chamber design. Wicks can be used to increase the area of the volatile
liquid:carrier gas interface but their presence, size and complexity is
limited by the internal resistance to gas flow created. The need for saturated
vapour output is balanced against the resistance created, and is simply
not achievable in all possible working conditions in draw over mode, particularly
at extremes of tidal volume or in cold environments.
Plenum vaporisers, with their constant driving
pressure and predictable flow rates can afford increased internal complexity
and resistance. Modern plenum vaporisers still have performance limitations
at extremes of flow rate and temperature, but they are generally more
accurate than their draw-over counterparts.
As vapour is liberated the temperature of the
liquid volatile agent falls due to the latent heat of vaporisation. This
causes a fall in the saturated vapour pressure and lowers the output of
the vaporiser. Temperature compensation is managed in two basic ways.
The first is to provide a large heat-sink of conductive material (water
bath or mass of metal), the dimensions of which are limited by size and
portability. Heat is conducted from the heat-sink to the volatile liquid
and minimises the fall in temperature of the liquid agent. The second
method is to vary the vapour chamber output with temperature, so that
more carrier gas is allowed to pass through the vapour chamber as the
temperature falls, and less as it rises. This is achieved by bimetallic
strips and ether-filled bellows in plenum vaporisers, but they cause an
increase in the internal resistance. Some drawover vaporisers have basic
thermo-compensation devices incorporated (EMO, PAC). In clinical practice
a fall in vaporiser output may be compensated for by an increased dial
setting.
Drawover vaporisers theoretically should not
be used in plenum fashion, as the output may not reflect the dialled setting.
This problem is more significant with some vaporisers, and is greatly
influenced by both flow rate and temperature. This is considered under
each vaporiser heading.
Most plenum vaporisers cannot be used for drawover
anaesthesia because of their high internal resistance.
EMO (Epstein Macintosh Oxford; Penlon; figure
2) is a classic design, unmodified since the 1950’s, which testifies
to its design and capabilities 8
. It is designed for use with ether and
is damaged by halothane. Stripping and maintenance is straightforward
(figure3). A key component is the temperature compensating device, which
is a sealed cannister containing liquid ether attached to a spindle, automated
by opposing springs. The splitting system comprises two concentric brass
cylinders with apertures, one of which rotates with the dial setter, thus
altering the overall ratio between vapour chamber and bypass flow. An
expensive setting gauge is available from Penlon to position the splitting
device correctly. A 0.1 inch (2.6mm, 8 French gauge, 12 Stubs needle gauge)
wire is an approximate substitute. To calibrate the dial properly one
must loosen the central screw, and place the dial in the 6% position.
The setting gauge is placed in the aperture, through the temperature compensator
portal, and the screw is tightened until the gauge is lightly gripped.
The vaporising chamber sits in a water bath, which acts as a heat sink.
This can be emptied for transport 8
. The entire EMO set-up weighs over 10kg,
limiting its potential for field use. In plenum mode the EMO only begins
to perform reasonably accurately with flow rates around 10 l/min, and
is therefore not ideal for paediatric use with a T piece, although circuit
adaptations can be made 9,10
.
If used in “pushover’ fashion,
with a ventilator or bellows placed upstream, the output can significantly
exceed the dial setting.
OMV (Oxford Miniature Vaporiser; Penlon; figure
4). This vaporiser is the most portable and most versatile drawover vaporiser,
but its size does impose performance limitations. The vapour chamber,
which contains 50mls of volatile agent, empties quite rapidly when in
use. It is suitable for a number of agents, a feature assisted by interchangeable
dial scales, and has basic thermal buffering in the form of a small glycol
(anti-freeze) reservoir within a metal heat sink 7,
11 . It suffers a reduction
in vapour output at lower temperatures, with a maximum output varying
from 2-4% with halothane between 0-30 O
C, and higher above this. Made from stainless
steel, it is resistant to corrosion by volatile agents. Metal mesh wicks
increase the output without significantly increasing the internal resistance.
The unit needs little regular maintenance. A common problem encountered
with the OMV is that the dial becomes stiff from thymol being deposited
in the mechanism during use with halothane. Thymol may be dissolved by
putting ether in the OMV and shaking it whilst working the lever back
and forward - remember to empty the unit afterwards! Alternatively strip
and clean the mechanism (if you have been trained).
It is common to use 2 OMV’s in series
to augment the output, as is standard in the Triservice apparatus, which
was originally used with trichloroethylene in one and halothane in the
other. The standard field anaesthetic machine of the Australian Defence
Forces uses two Oxford Miniature Vaporisers (OMV’s) in series in
either draw over or in true plenum mode (i.e. with continuous flow gases
fed in upstream), depending on the circuit attached.
It can operate efficiently as a plenum vaporiser
in anaesthesia and ICU 12,13
. Output reflects dial settings at 25
o C,
in either continuous flow or draw over use, but falls dramatically at
150 o C,
and rises steeply when above 35 o
C 14
. The reduction in output associated with
the fall in vaporiser temperature during use may require an increase in
dial setting, as determined clinically. Keeping the vaporiser topped up
with fresh liquid at room temperature helps maintain the output.
The OMV is reasonably accurate over a wide range
of flow rates and tidal volumes and, in particular it performs well at
small tidal volumes, making it suitable for paediatric anaesthesia
14,15 .
The OMV has also been used in a circle system.
However due to its efficiency it is capable of producing very high concentrations
and is not recommended for this use.
PAC (Portable Anaesthesia Complete; Datex-Ohmeda.
Now called TEC). Originally released as a series of individual vaporisers
designed for specific volatile agents 16
. A multi-agent version, the Ohmeda Universal
PAC, is now also available suitable for use with halothane, isoflurane,
enflurane, and diethyl ether. The apparent intention was to manufacture
the draw-over vaporiser with the best (linear) output performance profile
over a wide range of conditions, and that task has been achieved in adult
use. Accuracy is enhanced by a bimetallic strip temperature compensating
device, and there is a built in T-piece for oxygen supplementation. Unfortunately
the output is less accurate at small tidal volumes, or when used as a
plenum vaporizer with gas flows below 2-4l/min. Therefore it is not as
useful for paediatric anaesthesia. 14,16
It comes in a sturdy carrying case, and
has for many years been the standard issue field vaporiser used by the
US military and has been very widely used in Malawi 17,18
. In summary it is an excellent vaporiser,
particularly for adult draw over use. Regular servicing is recommended.
Self Inflating Bags/Bellows
Oxford Inflating Bellows (OIB) come as standard
with the EMO system (figure 2). The bellows sit vertically with a residual
internal volume maintained by a spring. This arrangement allows movement
of the bellows during spontaneous respiration providing a useful indicator
of breathing. The OIB was originally designed for use with a simple spring
loaded valve (eg Heidbrink valve). To facilitate gas flow through the
OIB there are two oneway valves in the form of metal discs on circular
seats. This arrangement works well for spontaneous ventilation (SV), but
is less than satisfactory for intermittent positive pressure ventilation
(IPPV) as adjustment of the Heidbrink valve must be constantly revised.
Non-rebreathing valves of either the Laerdal or Ambu type (figure 5) can
be used more effectively at the patient end of the draw over circuit to
facilitate IPPV, and are equally suitable for SV.
One note of caution is that with this adaptation
the OIB is prone to jam unless the downstream valve on the OIB is disabled
with the magnet provided (figure 6). When the OIB jams the patient cannot
exhale as an air-lock develops between the non-rebreathing valve and the
OIB valve. The patient must be disconnected from the circuit to allow
exhalation. This problem is more common with IPPV, but may also occur
in SV use. When in use the magnet holds the distal OIB flap valve in the
open position and stops the air-lock developing. Some anaesthetists even
remove the downstream disc to prevent this problem. A simpler, single
flap valve bellows called the Penlon Bellows Unit, PBU, has been developed
to address this issue, and to avoid confusion concerning when the magnet
should, and should not be used. If in doubt, it is useful to remember
that when using MODERN valves, use a MAGNET.
The tap on the side of the OIB is intended for
connection to supplemental oxygen when using the bellows for resuscitation
purposes. During anaesthesia, however, it is preferable to leave this
closed and supply oxygen upstream of the vaporiser. Adding oxygen at the
bellows dilutes the anaesthetic vapour.
To operate the bellows to assist ventilation
the recommended manoeuvre is a rocking motion, rather than direct up and
down. This creates less fatigue over time, and produces less variability
in tidal volume. The movement of the bellows during IPPV is characterised
by three phases: down, up, pause.
Laerdal, Ambu or other self inflating bags are
considered together as there is little practical difference between them.
Their valves are used to create the one-way flow in the circuit, and are
attached to the patient’s airway to minimise rebreathing. The bag
can be separated from the valve by a length of 22mm anaesthetic tubing
to allow it to sit better. The inflow side of the bag needs to be arranged
so that all gas is drawn through the vaporiser, and no air entrainment
is allowed, which would dilute the anaesthetic and potentially lead to
awareness. Spontaneous ventilation does not cause a movement of the bag
unless there is a fault in the draw over circuit causing a resistance
upstream. To observe gas flow in the circuit, tape a fine feather or piece
of paper at the inflow end of the whole system.
One Way Valves
The non-rebreathing valve (usually an Ambu valve
or Laerdal valve) should be placed as close to the patient’s airway
as possible to minimise the apparatus dead-space. Both valves can be scavenged.
A Heidbrink valve, or similar spring loaded blow-off valve, (which is
not one-way) can be used downstream of an Oxford Inflating Bellows during
spontaneous respiration, providing both OIB valves are functional. The
magnet should not be used to disable the valve in this instance.
Connecting tubing
The connecting tubing of a draw over circuit
is standard 22mm tubing. Antistatic tubing is required for ether, otherwise
lightweight plastic tubing is more convenient.
Conducting A Draw-Over Anaesthetic
Intravenous induction
This is performed as normal and the airway maintained
in an appropriate fashion. Face masks, endotracheal tubes and laryngeal
mask airways are all suitable for drawover anaesthesia.
Inhalational induction
During an inhalational (gaseous) induction a
seal is required between the face and the mask, or gas will not be drawn
through the vaporiser. When this occurs the patient will breathe room
air around the mask and remain conscious!
In adult anaesthesia it is relatively easy to
coax a mask on the face and still keep the patient calm and cooperative.
A benzodiazepine and/or opioid premedication will assist with this in
anxious patients, and markedly improves the tolerance to pungent volatile
agents such as ether or isoflurane. Problems may arise with elderly, edentulous
patients in whom masks may not fit well, or males with heavy beards in
whom the seal is hard to maintain. Filling the beard with lubricant jelly
does help, but makes the mask very slippery. A defibrillator pad or transparent
sticky plastic dressing with a hole cut in it fulfils the same need and
is easier to hold.
In paediatric use the problem is two-fold. The
child may be uncooperative so that maintaining a mask seal is difficult
(and sometimes psychologically traumatic). Small children (<15kg /
3 years) may not generate sufficient tidal volumes to draw vapour into
the circuit through the one-way valves, so even the cooperative ones may
be slow to induce!
One solution is to enlist an assistant to operate
the bellows (or self-inflating bag) to prime the circuit and bring vapour
up to the mask. Continued operation of the bellows will create flow through
the circuit and keep the supply of vapour coming, and the induction can
be done as if using a plenum anaesthetic system. The mask seal will not
be as important.
Adaptation of the drawover system to use with
a standard Ayre’s T-piece circuit is also possible for the very
young. This is done by connecting the T piece to the outlet of the OIB
19,20 .
The fresh gas flow is provided by the assistant slowly operating the bellows
6 - 8 times a minute and the T piece is used in the normal fashion.
The distal valve should be operational for this
system to be used. Paediatric drawover techniques are not detailed in
this review.
Maintenance of anaesthesia
Spontaneous ventilation has several advantages,
especially when using new or unfamiliar equipment and leaves the anaesthetist’s
hands free for other tasks. The volatile agent has to be delivered at
higher values to compensate for the absence of nitrous oxide. Parenteral
opioid analgesics should be provided, again titrated to signs of anaesthetic
depth. Alternatively supplementation by regional anaesthesia may be effective.
Neuromuscular blockade requires the patient to
be ventilated. This can be done manually, or through the use of a suitable
‘draw- over ventilator’, such as the Manley Multivent (Penlon
UK; figure 7).
Volatile Agents ( See
also Update in Anaesthesia No 11 - Volatile Anaesthetic Agents
).
Ether is still available in many parts of the
developing world, and where medical supplies are restricted industrial
grade ether may be successfully used. Unfortunately it is flammable in
air, and explosive in oxygen and can be ignited within a 25cm radius of
a source of ether vapour “zone of risk”. Provided sensible
antistatic precautions are taken to prevent any sources of sparking /
ignition within this zone ether may be safely used. During ether anaesthesia
diathermy should not be used in the airway, chest or upper abdomen. Ether
possesses excellent anaesthetic and analgesic properties. It has a low
potency which results in a prolonged inhalational induction with a well
described excitement phase. For this reason ether anaesthesia is easiest
provided following an intravenous induction. The OMV was originally designed
to assist with ether induction by additional use of halothane. It is placed
downstream of the EMO, and an OMV setting of 1% greatly speeds induction,
taking less than 10 minutes. The halothane can be turned off when the
ether concentration reaches 12 -15%.
Halothane (Fluothane) is widely available and
relatively inexpensive. It ranks highly as the agent of choice for use
in children. Although it slowly degrades some metals in the anaesthetic
equipment, and absorbs into rubber components of the circuit, it is an
excellent agent that has been used for nearly 50 years. Halothane contains
thymol which coats moving parts in the splitting system, and builds up
on wick devices, but can be removed with application of ether - see under
OMV.
Trichlorethylene (Trilene) has a relatively low
anaesthetic potency but provides good analgesia. In the Triservice apparatus
(Penlon) trichlorethylene and halothane were traditionally used in series.
It is becoming harder to source, although chemical reagent trichlorethyelene
is sometimes available.
Enflurane (Ethrane) is an agent that has rapidly
been phased out in many parts of the world after brief popularity in the
1980’s. It is still in use where economic factors favour it. It
can be used for inhalational induction and maintenance. Its main disadvantage
is its propensity to induce epileptiform phenomena, particularly with
hypocarbia and in children. Two OMV vaporisers are required to provide
adequate concentrations for induction. 21
Isoflurane (Forane) has the same saturated vapour
pressure as halothane, and is thus theoretically suitable for use in any
vaporiser designed for halothane. 22
Gaseous induction with isoflurane is
impaired by its relative pungency and airway irritative properties. With
benzodiazepine and / or opioid premedication, and a gentle approach, this
is an easily surmountable problem. Isoflurane has come down in price since
its patent expired.
Sevoflurane (Sevorane) has been used in draw-over,
but its use is hampered by a need to deliver high percentages which are
at the upper limits of simple vaporiser performance capabilities, as well
as its high cost. Using additional wicks to maximise output can be helpful,
but latent heat of vaporisation rapidly cools the system and lowers performance.
Two OMV vaporisers are required to provide adequate concentrations for
induction. 22
Right agent, wrong vaporiser?
Some vaporisers (OMV and TEC) are designed for
use with different agents and a variety of scales are provided to allow
this. Care should be taken that agents (eg ether, trichlorethylene and
halothane) are not mistakenly put in the wrong vaporiser, or the wrong
dial used.
Conclusion
Draw-over anaesthesia holds distinct advantages
in less affluent hospitals where the advantages over any other form of
anaesthetic delivery system are not only economic, but also practical,
and the training required to deliver safe anaesthesia is kept to a minimum.
During field anaesthesia the added attractions of portability and reliability
allows rapid and effective deployment of anaesthesia equipment to the
area of need. Field anaesthesia, such as that performed by the military
and humanitarian organisations, is greatly simplified by combining draw-over
and intravenous techniques. Major surgery is made possible that is undoubtedly
life saving. In the first world hospital, exposure of trainees to draw-over
techniques results in a deeper understanding of equipment and more skilled
anaesthetist.
Our thanks for the constructive and expert assistance
with this article given by Dr Haydn Perndt, Staff Anaesthetist and Course
Director of “Remote Situations, Difficult Circumstances, and Developing
Country Anaesthesia” Course, Royal Hobart Hospital, Tasmania, Australia.
References
1. Akinyemi OO, Adelaja AB. Blood gas studies
using spontaneously respired halothane in ambient air. Anaesthesia
1982; 36
:353-4
2. Tighe SQ, Turner GA, Merrill SB, Pethybridge
RJ. Minimum oxygen requirements during anaesthesia with the Triservice
anaesthetic apparatus. A study of drawover anaesthesia in the young adult.
Anaesthesia 1991;
46 :52-6
3. Mackie AM. Drawover anaesthetic systems. Factors
determining the inspired oxygen concentration. Anaesthesia
1987 ; 42
:299-304
4. Wilson IH, van Heerden PV, Leigh J. Domiciliary
oxygen concentrators in anaesthesia: preoxygenation techniques and inspired
oxygen concentrations. British Journal
of Anaesthesia 1990; 65
:342-5
5. Dobson MB. Oxygen concentrators for the smaller
hospital-a review. Tropical Doctor
1992; 22
:56-8
6. Dobson MB. Oxygen concentrators for district
hospitals. Update in Anaesthesia
1999;10:
7. Houghton IT. The Triservice anaesthetic apparatus.
Anaesthesia 1981;
36 :1094-108
8. Ball C, Westhorpe R. The EMO vaporizer.
Anaesthesia & Intensive Care
1998; 26
:347
9. Schaefer HG, Farman JV. Anaesthetic vapour
concentrations in the EMO system. Anaesthesia
1984; 39
:171-80
10. Marsh DR, Herbert P. Performance of the EMO
inhaler. Anaesthesia .
38(6):575-7, 1983
11. Page RJE, Wilson IH. Drawover anaesthesia.
British Journal of Hospital Medicine
1989; 42
:320-2
12. McIndoe AK, Stewart P, Wilson IH. Drawover
vaporizers for sedation in intensive care.
Intensive Care Medicine 1997;
23 :704-7
13. Taylor JC, Restall J. Can a drawover vaporizer
be a pushover? Anaesthesia
1994; 49
:892-4
14. Craig GR, Berry CB, Yeats MJ. An evaluation
of the Universal PAC and Oxford Miniature Vaporizers for paediatric field
anaesthesia. Anaesthesia
1995; 50
:789-93
15. Wilson IH, Page RJE, Yeats MJ. The Oxford
Miniature Vaporizer in paediatric anaesthesia. An experimental study.
Anaesthesia 1988;
43 :700-2
16. Borland CW, et al. Evaluation of a new range
of air drawover vaporisers. The PAC series - laboratory and field studies.
Anaesthesia 1983;
38 ;852-862.
17. Fenton PM. The Malawi anaesthetic machine.
Experience with a new type of anaesthetic apparatus for developing countries.
Anaesthesia .
1989; 44 :498-503
18. Pedersen J, Nyrop M. Anaesthetic equipment
for a developing country. British Journal
of Anaesthesia . 1991; 66
:264-70
19. Bewes P. Anaesthesia in children using the
EMO system. Update in Anaesthesia No.8
.
20. Hodges S Letter regarding Anaesthesia in
children using the EMO system. Update
in Anaesthesia 1998; 9
:52
21. M Kocan. The Triservice anaesthetic apparatus.
Trial of isoflurane and enflurane as alternatives to halothane.
Anaesthesia
1987; 42 :1101-
4.
22. Liu EH. Dhara SS. Sevoflurane output from
the Oxford Miniature Vaporizer in drawover mode. Anaesthesia
& Intensive Care . 2000;
28 :532-6
|